Monopolar Forceps and Scissors
Used for used for tissue grasping, cutting, and coagulation,
These devices are reusable and are supplied non-sterile.
Process through cleaning and sterilization prior to initial use.
We guarantee our products to withstand a minimum of 20 sterilization cycles
Care in use and handling can extend useful life.
Sterilization
Wrap each pair of instrument separately or place in a container so as to prevent instrument from contacting
each other or other instruments.
Steam: Steam autoclave at 250°F (12ºC) for 30 minutes.
Flash: Steam autoclave at 275°F (134ºC) for 10 minutes.
Pre-Vac: Steam autoclave at 270°-273°F (132°- 135°C), Exposure Time (4 minutes), Dry Time (20 minutes)
These devices are reusable and are supplied non-sterile.
Process through cleaning and sterilization prior to initial use.
We guarantee our products to withstand a minimum of 20 sterilization cycles
Care in use and handling can extend useful life.
Sterilization
Wrap each pair of instrument separately or place in a container so as to prevent instrument from contacting
each other or other instruments.
Steam: Steam autoclave at 250°F (12ºC) for 30 minutes.
Flash: Steam autoclave at 275°F (134ºC) for 10 minutes.
Pre-Vac: Steam autoclave at 270°-273°F (132°- 135°C), Exposure Time (4 minutes), Dry Time (20 minutes)